1.Imported grating as dispersive element to change wavelength
2.Light source: Halogen-tungsten lamp (12 V20W)
3.Wavelength range (nm): 330---800nm
4.Wavelength accuracy: ±2nm
5.Wavelength reproducibility: <=1nm
6.Bandwidth: <=6nm
7.Absorbance range: -0.3Abs~3.0Abs
8.Linear error: ±5%
9.Absorbance reproducibility: <=0.005Abs
10.Resolving capability: 0.001Abs
11.Temperature control: room temperature, 25°C, 30°C or 37°C by Peltier device
12.Temperature-control accuracy: ± 0.5°C
13.Pump volume: 200ul~2000ul
14.Testing method:
(1) Linear end point
(2) Non-linear end point (standard curve)
(3) Two-point (time fixed)
(4) Kinetic
(5) Double wavelength
(6) Absorbance
(7) Sample blank
15.Items: 72 fixed items, four free positions for up to 200 self-defined items
16.Keyboard: membrane keyboard
17.Display: 10.4” color LCD
18.Printer: built-in thermal printer, paper width: 110mm
19.Environmental temperature: 10°C~30°C
20.Relative environmental humidity: <=80%
21.Drift: <=0.005Abs/30min
22.Cell: quartz flowing; 32ul
23.Cross contamination: <=2%
24.Open program
25.Reagent: open, free choice
26.Standard RS-232 interface
27.Power supply: 220V±22 V, 50Hz ±1Hz
28.Size(mm): 320×340×440
29.Weight: 12.5Kg
Structure:
Power switch, monochromator, flowcell, temperature-control system, sample-filling system, printer, microcomputer system, color LCD, keyboard.
Features:
(1) Plane holographic diffraction grating as dispersive element.
(2) color LCD, easy operation.
(3) built-in micro-computer to control and to handle the test.
(4) auto printing, auto memory, auto wavelength change, real-time display.
(5) testing method: Linear end point,non-linear end point (standard curve), two-point (fixed time), kinetic, double wavelength, absorbance, sample blank
(6) open programming for all items, agent free-choice.
(7) 72 fixed items, 4 free positions for up to 200 self-defined items.
(8) data protection from power off.
(9) fixed factor method for items with standard, once-only targeting.
(10) print sample report, item report and parameter chart for each programmed item.
(11) auto storing QC within a month, QC chart and statistics displayed and printed available.
(12) standard curve and response linearity displayed, results detected.
(13) recalculating and reprinting results according to QC statistics.
(14) time setup, pump volume calibration, printing color adjusting, breakdowns warning, protection from power-off.
(15) RS-232 interface to send data to computer.
Big Screen Semi-auto Biochemistry Analyzer